Validation of an Accelerometer Based BCG Method for Sleep Analysis

Loading...
Thumbnail Image

URL

Journal Title

Journal ISSN

Volume Title

School of Electrical Engineering | D4 Julkaistu kehittämis- tai tutkimusraportti tai -selvitys

Date

2016

Major/Subject

Mcode

Degree programme

Language

en

Pages

12

Series

Aalto University publication series SCIENCE + TECHNOLOGY, 7/2016

Abstract

Sleep problems are one of the most common medical complaints today. Polysomnography (PSG) as the current standard for sleep analysis is expensive, intrusive and complex. Thus, finding a reliable and unobtrusive method for longer-term home use is important. Ballistocardiography (BCG) based methods have shown potential in sleep analysis recently. The usability and performance of a BCG based method in qualitative and quantitative analysis of sleep was evaluated. The method was validated in a clinical test on 20 subjects using PSG as a reference. Heart rate (HR), heart rate variability (HRV), respiratory rate (RR), respiratory rate variability (RRV), respiratory depth (Rdepth) and movement were utilized for sleep stage detection. The BCG parameter accuracy was presented as the mean error from PSG with 95% confidence interval. The errors were -0.1 ± 4.4 beats per minute for HR, -0.9 ± 14.7 ms for high frequency (HF) HRV, -3.0 ± 29.9 ms for low frequency (LF) HRV, 0.3 ± 4.5 breaths per minute for RR and -40 ± 424 ms for RRV respectively. Correlation coefficient was 0.97 for HR, 0.67 for HF HRV, 0.71 for LF HRV, 0.54 for RR and 0.49 for RRV. HR, RRV and Rdepth were typically at an increased level in REM sleep and wakefulness and decreased in deep sleep. RRV was at its highest during wakefulness. HRV was at a decreased level in REM and wakefulness and increased in deep sleep. Movement was higher during wakefulness than in sleep.

Description

Keywords

ballistocardiography, polysomnography, sleep quality

Other note

Citation