Fast Electron Thermometry for Ultrasensitive Calorimetric Detection
| dc.contributor | Aalto-yliopisto | fi |
| dc.contributor | Aalto University | en |
| dc.contributor.author | Gasparinetti, S. | |
| dc.contributor.author | Viisanen, Klaara | |
| dc.contributor.author | Saira, Olli-Pentti | |
| dc.contributor.author | Faivre, T. | |
| dc.contributor.author | Arzeo, M. | |
| dc.contributor.author | Meschke, M. | |
| dc.contributor.author | Pekola, J.P. | |
| dc.contributor.department | Department of Applied Physics | en |
| dc.contributor.groupauthor | Quantum Phenomena and Devices | en |
| dc.date.accessioned | 2017-06-20T11:20:00Z | |
| dc.date.available | 2017-06-20T11:20:00Z | |
| dc.date.issued | 2015 | |
| dc.description | VK: Low Temperature Laboratory | |
| dc.description.abstract | We demonstrate radio-frequency thermometry on a micrometer-sized metallic island below 100 mK. Our device is based on a normal-metal–insulator–superconductor tunnel junction coupled to a resonator with transmission readout. In the first generation of the device, we achieve 90 μK/√Hz noise-equivalent temperature with 10 MHz bandwidth. We measure the thermal relaxation time of the electron gas in the island, which we find to be of the order of 100 μs. Such a calorimetric detector, upon optimization, can be seamlessly integrated into superconducting circuits, with immediate applications in quantum-thermodynamics experiments down to single quanta of energy. | en |
| dc.description.version | Peer reviewed | en |
| dc.format.mimetype | application/pdf | |
| dc.identifier.citation | Gasparinetti, S, Viisanen, K, Saira, O-P, Faivre, T, Arzeo, M, Meschke, M & Pekola, J P 2015, 'Fast Electron Thermometry for Ultrasensitive Calorimetric Detection', Physical Review Applied, vol. 3, no. 1, 014007, pp. 1-7. https://doi.org/10.1103/PhysRevApplied.3.014007 | en |
| dc.identifier.doi | 10.1103/PhysRevApplied.3.014007 | |
| dc.identifier.issn | 2331-7019 | |
| dc.identifier.other | PURE UUID: acb69b6f-e59c-498d-8f6e-2c4e9cc6ab12 | |
| dc.identifier.other | PURE ITEMURL: https://research.aalto.fi/en/publications/acb69b6f-e59c-498d-8f6e-2c4e9cc6ab12 | |
| dc.identifier.other | PURE LINK: http://dx.doi.org/10.1103/PhysRevApplied.3.014007 | |
| dc.identifier.other | PURE FILEURL: https://research.aalto.fi/files/13439400/PhysRevApplied.3.014007.pdf | |
| dc.identifier.uri | https://aaltodoc.aalto.fi/handle/123456789/26981 | |
| dc.identifier.urn | URN:NBN:fi:aalto-201706205705 | |
| dc.language.iso | en | en |
| dc.publisher | American Physical Society | |
| dc.relation.ispartofseries | Physical Review Applied | en |
| dc.relation.ispartofseries | Volume 3, issue 1, pp. 1-7 | en |
| dc.rights | openAccess | en |
| dc.title | Fast Electron Thermometry for Ultrasensitive Calorimetric Detection | en |
| dc.type | A1 Alkuperäisartikkeli tieteellisessä aikakauslehdessä | fi |
| dc.type.version | publishedVersion |
Files
Original bundle
1 - 1 of 1
Loading...
- Name:
- PhysRevApplied.3.014007.pdf
- Size:
- 556.79 KB
- Format:
- Adobe Portable Document Format