Flat-field calibration method for hyperspectral frame cameras
Loading...
Access rights
openAccess
publishedVersion
URL
Journal Title
Journal ISSN
Volume Title
A1 Alkuperäisartikkeli tieteellisessä aikakauslehdessä
This publication is imported from Aalto University research portal.
View publication in the Research portal (opens in new window)
View/Open full text file from the Research portal (opens in new window)
View publication in the Research portal (opens in new window)
View/Open full text file from the Research portal (opens in new window)
Unless otherwise stated, all rights belong to the author. You may download, display and print this publication for Your own personal use. Commercial use is prohibited.
Date
Major/Subject
Mcode
Degree programme
Language
en
Pages
8
Series
Metrologia
Abstract
This paper presents a method for characterising spatial responsivity of hyperspectral cameras. Knowing the responsivity of the camera as a function of pixel coordinates allows applying a flat-field correction on image data. The method is based on scanning the field of view of the camera with a broadband radiance source, based on an integrating sphere, and combining the captured frames to synthesise a uniform radiance source filling the whole field of view of the camera at the focus distance. The method was compared with a traditional approach where the aperture of an integrating sphere is imaged from a close distance, filling the entire field of view of the camera. The measurement setup was tested with a hyperspectral camera, based on a tunable Fabry–Pérot interferometer. Without the flat-field correction, the average standard deviation of the pixel responsivities across all the spectral channels of the camera was 3.78%. After the correction, the average standard deviation was reduced to 0.40% and 3.25% for the aperture-scanning method and the close-distance method, respectively. The expanded uncertainty (k = 2) for the flat-field correction obtained using the scanning method was 0.68%–0.78%, depending on the spectral channel of the camera.
Description
Other note
Citation
Kokka, A, Pulli, T, Honkavaara, E, Markelin, L, Kärhä, P & Ikonen, E 2019, 'Flat-field calibration method for hyperspectral frame cameras', Metrologia. https://doi.org/10.1088/1681-7575/ab3261