dc.description.abstractWe demonstrate successful “dry” refrigeration of quantum fluids down to T = 0.16 mK by using copper nuclear demagnetization stage that is pre-cooled by a pulse-tube-based dilution refrigerator. This type of refrigeration delivers a flexible and simple sub-mK solution to a variety of needs including experiments with superfluid 3He. Our central design principle was to eliminate relative vibrations between the high-field magnet and the nuclear refrigeration stage, which resulted in the minimum heat leak of Q = 4.4 nW obtained in field of 35 mT. For thermometry, we employed a quartz tuning fork immersed into liquid 3He. We show that the fork oscillator can be considered as self-calibrating in superfluid 3He at the crossover point from hydrodynamic into ballistic quasiparticle regime.en
dc.description.versionPeer revieweden
dc.format.extent085106/1-6
dc.format.mimetypeapplication/pdfen
dc.identifier.citationTodoshchenko, I. & Kaikkonen, J.-P. & Blaauwgeers, R. & Hakonen, Pertti J. & Savin, A. 2014. Dry demagnetization cryostat for sub-millikelvin helium experiments: Refrigeration and thermometry. Review of Scientific Instruments. Volume 85, Issue 8. 085106/1-6. ISSN 0034-6748 (printed). DOI: 10.1063/1.4891619en
dc.identifier.doi10.1063/1.4891619
dc.identifier.issn0034-6748 (printed)
dc.identifier.urihttps://aaltodoc.aalto.fi/handle/123456789/17876
dc.identifier.urnURN:NBN:fi:aalto-201509254470
dc.language.isoenen
dc.publisherAIP Publishingen
dc.relation.ispartofseriesReview of Scientific Instrumentsen
dc.relation.ispartofseriesVolume 85, Issue 8
dc.rights© 2014 AIP Publishing. This article may be downloaded for personal use only. Any other use requires prior permission of the authors and the American Institute of Physics. The following article appeared in Review of Scientific Instruments, Volume 85, Issue 8 and may be found at http://scitation.aip.org/content/aip/journal/rsi/85/8/10.1063/1.4891619.en
dc.rights.holderAIP Publishing
dc.subject.keywordheliumen
dc.subject.keyworddemagnetizationen
dc.subject.keywordcopperen
dc.subject.keywordrefrigeratorsen
dc.subject.keywordthermometersen
dc.subject.otherPhysicsen
dc.titleDry demagnetization cryostat for sub-millikelvin helium experiments: Refrigeration and thermometryen
dc.typeA1 Alkuperäisartikkeli tieteellisessä aikakauslehdessäfi
dc.type.dcmitypetexten
dc.type.versionFinal published versionen

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
A1_todoshchenko_i_2014.pdf
Size:
987.72 KB
Format:
Adobe Portable Document Format