Learning Centre

Effect of serum dilution fluids on the wear of unirradiated and high dose gamma-irradiated, vitamin E stabilized UHMWPE

 |  Login

Show simple item record

dc.contributor Aalto-yliopisto fi
dc.contributor Aalto University en
dc.contributor.author Saikko, Vesa
dc.date.accessioned 2019-09-03T13:48:02Z
dc.date.available 2019-09-03T13:48:02Z
dc.date.issued 2019-07-15
dc.identifier.citation Saikko , V 2019 , ' Effect of serum dilution fluids on the wear of unirradiated and high dose gamma-irradiated, vitamin E stabilized UHMWPE ' , Wear , vol. 430-431 , pp. 76-80 . https://doi.org/10.1016/j.wear.2019.04.022 en
dc.identifier.issn 0043-1648
dc.identifier.issn 1873-2577
dc.identifier.other PURE UUID: aa48e854-0792-4665-9adc-ace6d41dcd77
dc.identifier.other PURE ITEMURL: https://research.aalto.fi/en/publications/aa48e854-0792-4665-9adc-ace6d41dcd77
dc.identifier.other PURE LINK: http://www.scopus.com/inward/record.url?scp=85064634757&partnerID=8YFLogxK
dc.identifier.other PURE FILEURL: https://research.aalto.fi/files/36337658/ENG_Saikko_Vesa_Effect_of_serum_dilution_fluids_Wear.pdf
dc.identifier.uri https://aaltodoc.aalto.fi/handle/123456789/40100
dc.description.abstract In orthopaedic tribology, the ISO 14242-1 standard recommends that the serum lubricant is diluted with deionized water (DW). Sometimes, phosphate buffered saline (PBS) is used instead for the dilution. Noncyclic, multidirectional pin-on-disk tests showed that the dilution of alpha calf serum with PBS or phosphate buffer resulted in a 5–40-fold decrease in the wear factor of unirradiated and of vitamin E stabilized, highly crosslinked ultrahigh molecular weight polyethylene (UHMWPE) compared with DW dilution. It appears that the phosphates were mainly responsible for this phenomenon. PBS dilution may result in a substantial underestimation of clinical wear. With 1:1 DW dilution, the mean wear factor of highly crosslinked UHMWPE was 5.7-fold lower than that of unirradiated UHMWPE. en
dc.format.extent 5
dc.format.extent 76-80
dc.format.mimetype application/pdf
dc.language.iso en en
dc.publisher Elsevier BV
dc.relation.ispartofseries Wear en
dc.relation.ispartofseries Volume 430-431 en
dc.rights openAccess en
dc.title Effect of serum dilution fluids on the wear of unirradiated and high dose gamma-irradiated, vitamin E stabilized UHMWPE en
dc.type A1 Alkuperäisartikkeli tieteellisessä aikakauslehdessä fi
dc.description.version Peer reviewed en
dc.contributor.department Department of Mechanical Engineering
dc.subject.keyword Biomedical devices
dc.subject.keyword Lubrication environment
dc.subject.keyword Noncyclic pin-on-disk
dc.subject.keyword UHMWPE
dc.identifier.urn URN:NBN:fi:aalto-201909035142
dc.identifier.doi 10.1016/j.wear.2019.04.022
dc.date.embargo info:eu-repo/date/embargoEnd/2021-04-28