Title: | Safety and recommendations for TMS use in healthy subjects and patient populations, with updates on training, ethical and regulatory issues |
Author(s): | Rossi, Simone ; Antal, Andrea ; Bestmann, Sven ; Bikson, Marom ; Brewer, Carmen ; Brockmöller, Jürgen ; Carpenter, Linda L. ; Cincotta, Massimo ; Chen, Robert ; Daskalakis, Jeff D. ; Di Lazzaro, Vincenzo ; Fox, Michael D. ; George, Mark S. ; Gilbert, Donald ; Kimiskidis, Vasilios K. ; Koch, Giacomo ; Ilmoniemi, Risto J. ; Pascal Lefaucheur, Jean ; Leocani, Letizia ; Lisanby, Sarah H. ; Miniussi, Carlo ; Padberg, Frank ; Pascual-Leone, Alvaro ; Paulus, Walter ; Peterchev, Angel V. ; Quartarone, Angelo ; Rotenberg, Alexander ; Rothwell, John ; Rossini, Paolo M. ; Santarnecchi, Emiliano ; Shafi, Mouhsin M. ; Siebner, Hartwig R. ; Ugawa, Yoshikatzu ; Wassermann, Eric M. ; Zangen, Abraham ; Ziemann, Ulf ; Hallett, Mark |
Date: | 2021-01 |
Language: | en |
Pages: | 269–306 |
Department: | University of Siena University of Göttingen University College London City University of New York National Institutes of Health Brown University Central Tuscany Local Health Authority University of Toronto Universita Campus Bio-Medico di Roma Harvard Medical School Medical University of South Carolina Cincinnati Children's Hospital Medical Center Aristotle University of Thessaloniki IRCCS Fondazione Santa Lucia - Roma Department of Neuroscience and Biomedical Engineering Université Paris-Est Créteil Vita-Salute San Raffaele University BEC-INFM Ludwig Maximilian University of Munich Children's Healthcare of Atlanta Duke University University of Messina IRCCS San Raffaele Pisana - Roma University of Copenhagen Fukushima Medical University Ben-Gurion University of the Negev University of Tübingen |
Series: | Clinical Neurophysiology, Volume 132, issue 1 |
ISSN: | 1388-2457 1872-8952 |
DOI-number: | 10.1016/j.clinph.2020.10.003 |
Keywords: | Neurology, Neuromodulation, Psychiatry, QPS, rTMS, Safety, TBS, TMS |
|
|
This publication is imported from Aalto University Research information portal: https://research.aalto.fi
> View this publication in Research information portal > Other link related to publication (Research information portal) |
|
Rossi , S , Antal , A , Bestmann , S , Bikson , M , Brewer , C , Brockmöller , J , Carpenter , L L , Cincotta , M , Chen , R , Daskalakis , J D , Di Lazzaro , V , Fox , M D , George , M S , Gilbert , D , Kimiskidis , V K , Koch , G , Ilmoniemi , R J , Pascal Lefaucheur , J , Leocani , L , Lisanby , S H , Miniussi , C , Padberg , F , Pascual-Leone , A , Paulus , W , Peterchev , A V , Quartarone , A , Rotenberg , A , Rothwell , J , Rossini , P M , Santarnecchi , E , Shafi , M M , Siebner , H R , Ugawa , Y , Wassermann , E M , Zangen , A , Ziemann , U & Hallett , M 2021 , ' Safety and recommendations for TMS use in healthy subjects and patient populations, with updates on training, ethical and regulatory issues : Expert Guidelines ' , Clinical Neurophysiology , vol. 132 , no. 1 , pp. 269–306 . https://doi.org/10.1016/j.clinph.2020.10.003 |
|
Abstract:This article is based on a consensus conference, promoted and supported by the International Federation of Clinical Neurophysiology (IFCN), which took place in Siena (Italy) in October 2018. The meeting intended to update the ten-year-old safety guidelines for the application of transcranial magnetic stimulation (TMS) in research and clinical settings (Rossi et al., 2009). Therefore, only emerging and new issues are covered in detail, leaving still valid the 2009 recommendations regarding the description of conventional or patterned TMS protocols, the screening of subjects/patients, the need of neurophysiological monitoring for new protocols, the utilization of reference thresholds of stimulation, the managing of seizures and the list of minor side effects. New issues discussed in detail from the meeting up to April 2020 are safety issues of recently developed stimulation devices and pulse configurations; duties and responsibility of device makers; novel scenarios of TMS applications such as in the neuroimaging context or imaging-guided and robot-guided TMS; TMS interleaved with transcranial electrical stimulation; safety during paired associative stimulation interventions; and risks of using TMS to induce therapeutic seizures (magnetic seizure therapy). An update on the possible induction of seizures, theoretically the most serious risk of TMS, is provided. It has become apparent that such a risk is low, even in patients taking drugs acting on the central nervous system, at least with the use of traditional stimulation parameters and focal coils for which large data sets are available. Finally, new operational guidelines are provided for safety in planning future trials based on traditional and patterned TMS protocols, as well as a summary of the minimal training requirements for operators, and a note on ethics of neuroenhancement.
|
|
|
|
|